OEM#: 3481039New OEM Original Outright
IMPULSE 7000DP US 120V WITH IMPULSE 7010 AND TEST AUTOMATION by Fluke Electronics Corp (Biomedical Div.)
Your price:
$13,535.72 each

OEM#: 3481039IMPULSE 7000DP US 120V WITH IMPULSE 7010 AND TEST AUTOMATION by Fluke Electronics Corp (Biomedical Div.)
New OEM Original Outright
Your price:
$13,535.72 each
Asset Information
Are you an existing customer? Sign in to access your account.
ADD TO LIST
Product Overview
The Impulse 7000DP Defibrillator/Transcutaneous Pacemaker Analyzer ensures proper operation and ultimate performance of critical life-support cardiac-resuscitation equipment. The Impulse 7000DP is superior in compliance, technology, efficiency and ease of use. A sleek and rugged exterior is paired with first-class technology, putting this defib tester in a league of its own. Standardize test procedures, capture, and print results with Ansur test-automation software, or use the Impulse 7010 Defibrillator Selectable Load Accessory to ensure IEC 60601-2-4 and AAMI DF80 compliance.
Features
- First-in-class accuracy ± 1% of reading + 0.1J
- Efficient, accurate, versatile, trustworthy
- Long lasting, rechargeable battery
- Impulse 7010 Defibrillator Selectable Load Accessory provides multiple loads of 25 Ω, 50 Ω, 75 Ω, 100 Ω, 125 Ω, 150 Ω, 175 Ω, and 200 Ω to comply with IEC 60601-2-4 standard (optional)
- Lown, edmark, trapezoidal, biphasic and pulsed-biphasic defibrillation technology compatibility
- AED compatibility
- Ability to test external transcutaneous pacemakers (Impulse 7000 DP only)
- Internal pacer brand selections
- Flexible heart-rate settings (1BPM step)
- DSP-based measurements for firmware and waveform upgrade
- Waveform capture, store and replay
Technical Specs
P/N: 3481039 Key Specs
Item
IMPULSE 7000DP US 120V WITH IMPULSE 7010 AND TEST AUTOMATION
OEM
Fluke Electronics Corp (Biomedical Div.)
OEM #
3481039
Condition
New OEM Original
Returnable
No
Warranty
12 Month
Accuracy
±1 %, non-inductive (< 2 µH)
Auto-ranged Measurement
0.1 J to 600 J
Maximum Average Power
12 W, equivalent to 10 defib pulses of 360 J every 5 minutes
Pulse Trigger Level
20 V
Resistance
50 Ω
Sample Rate
250 kHz (4 µs sample)
Scope Output
Autorange:2000:1, 400:1 and 80:1: dependant on the range
Waveform Playback
Output: BNC; Output impedance: 50 Ω; Amplitude Accuracy: ± 5 %
Fitments
The Impulse 7000DP Defibrillator/Transcutaneous Pacemaker Analyzer ensures proper operation and ultimate performance of critical life-support cardiac-resuscitation equipment. The Impulse 7000DP is superior in compliance, technology, efficiency and ease of use. A sleek and rugged exterior is paired with first-class technology, putting this defib tester in a league of its own. Standardize test procedures, capture, and print results with Ansur test-automation software, or use the Impulse 7010 Defibrillator Selectable Load Accessory to ensure IEC 60601-2-4 and AAMI DF80 compliance.
- Fluke Electronics Corp (Biomedical Div.)
- IMPULSE 7000DP
- TA-IMP7K/7010 US
Product Overview
The Impulse 7000DP Defibrillator/Transcutaneous Pacemaker Analyzer ensures proper operation and ultimate performance of critical life-support cardiac-resuscitation equipment. The Impulse 7000DP is superior in compliance, technology, efficiency and ease of use. A sleek and rugged exterior is paired with first-class technology, putting this defib tester in a league of its own. Standardize test procedures, capture, and print results with Ansur test-automation software, or use the Impulse 7010 Defibrillator Selectable Load Accessory to ensure IEC 60601-2-4 and AAMI DF80 compliance.
Features
- First-in-class accuracy ± 1% of reading + 0.1J
- Efficient, accurate, versatile, trustworthy
- Long lasting, rechargeable battery
- Impulse 7010 Defibrillator Selectable Load Accessory provides multiple loads of 25 Ω, 50 Ω, 75 Ω, 100 Ω, 125 Ω, 150 Ω, 175 Ω, and 200 Ω to comply with IEC 60601-2-4 standard (optional)
- Lown, edmark, trapezoidal, biphasic and pulsed-biphasic defibrillation technology compatibility
- AED compatibility
- Ability to test external transcutaneous pacemakers (Impulse 7000 DP only)
- Internal pacer brand selections
- Flexible heart-rate settings (1BPM step)
- DSP-based measurements for firmware and waveform upgrade
- Waveform capture, store and replay
Technical Specs
P/N: 3481039 Key Specs
Item
IMPULSE 7000DP US 120V WITH IMPULSE 7010 AND TEST AUTOMATION
OEM
Fluke Electronics Corp (Biomedical Div.)
OEM #
3481039
Condition
New OEM Original
Returnable
No
Warranty
12 Month
Accuracy
±1 %, non-inductive (< 2 µH)
Auto-ranged Measurement
0.1 J to 600 J
Maximum Average Power
12 W, equivalent to 10 defib pulses of 360 J every 5 minutes
Pulse Trigger Level
20 V
Resistance
50 Ω
Sample Rate
250 kHz (4 µs sample)
Scope Output
Autorange:2000:1, 400:1 and 80:1: dependant on the range
Waveform Playback
Output: BNC; Output impedance: 50 Ω; Amplitude Accuracy: ± 5 %
Fitments
The Impulse 7000DP Defibrillator/Transcutaneous Pacemaker Analyzer ensures proper operation and ultimate performance of critical life-support cardiac-resuscitation equipment. The Impulse 7000DP is superior in compliance, technology, efficiency and ease of use. A sleek and rugged exterior is paired with first-class technology, putting this defib tester in a league of its own. Standardize test procedures, capture, and print results with Ansur test-automation software, or use the Impulse 7010 Defibrillator Selectable Load Accessory to ensure IEC 60601-2-4 and AAMI DF80 compliance.
- Fluke Electronics Corp (Biomedical Div.)
- IMPULSE 7000DP
- TA-IMP7K/7010 US
Product Overview
The Impulse 7000DP Defibrillator/Transcutaneous Pacemaker Analyzer ensures proper operation and ultimate performance of critical life-support cardiac-resuscitation equipment. The Impulse 7000DP is superior in compliance, technology, efficiency and ease of use. A sleek and rugged exterior is paired with first-class technology, putting this defib tester in a league of its own. Standardize test procedures, capture, and print results with Ansur test-automation software, or use the Impulse 7010 Defibrillator Selectable Load Accessory to ensure IEC 60601-2-4 and AAMI DF80 compliance.
Features
- First-in-class accuracy ± 1% of reading + 0.1J
- Efficient, accurate, versatile, trustworthy
- Long lasting, rechargeable battery
- Impulse 7010 Defibrillator Selectable Load Accessory provides multiple loads of 25 Ω, 50 Ω, 75 Ω, 100 Ω, 125 Ω, 150 Ω, 175 Ω, and 200 Ω to comply with IEC 60601-2-4 standard (optional)
- Lown, edmark, trapezoidal, biphasic and pulsed-biphasic defibrillation technology compatibility
- AED compatibility
- Ability to test external transcutaneous pacemakers (Impulse 7000 DP only)
- Internal pacer brand selections
- Flexible heart-rate settings (1BPM step)
- DSP-based measurements for firmware and waveform upgrade
- Waveform capture, store and replay
Technical Specs
P/N: 3481039 Key Specs
Item
IMPULSE 7000DP US 120V WITH IMPULSE 7010 AND TEST AUTOMATION
OEM
Fluke Electronics Corp (Biomedical Div.)
OEM #
3481039
Condition
New OEM Original
Returnable
No
Warranty
12 Month
Accuracy
±1 %, non-inductive (< 2 µH)
Auto-ranged Measurement
0.1 J to 600 J
Maximum Average Power
12 W, equivalent to 10 defib pulses of 360 J every 5 minutes
Pulse Trigger Level
20 V
Resistance
50 Ω
Sample Rate
250 kHz (4 µs sample)
Scope Output
Autorange:2000:1, 400:1 and 80:1: dependant on the range
Waveform Playback
Output: BNC; Output impedance: 50 Ω; Amplitude Accuracy: ± 5 %
Fitments
The Impulse 7000DP Defibrillator/Transcutaneous Pacemaker Analyzer ensures proper operation and ultimate performance of critical life-support cardiac-resuscitation equipment. The Impulse 7000DP is superior in compliance, technology, efficiency and ease of use. A sleek and rugged exterior is paired with first-class technology, putting this defib tester in a league of its own. Standardize test procedures, capture, and print results with Ansur test-automation software, or use the Impulse 7010 Defibrillator Selectable Load Accessory to ensure IEC 60601-2-4 and AAMI DF80 compliance.
- Fluke Electronics Corp (Biomedical Div.)
- IMPULSE 7000DP
- TA-IMP7K/7010 US